To obtain Elix® pure
water quality, the purification sequence has been fully optimized. The patented
Elix® module (EDI technology) is combined with Progard®complete
pretreatment, intelligent reverse osmosis and optional bactericidal UV lamp
technologies. Depending on daily volume needs, different production flow rates
are available (3, 5, 10 & 15 L/h). Only Elix® systems are equipped
with the best-in-class unique Elix® module. It provides a superior and
consistent water quality while preserving the environment (self-regeneration of
resin beads, minimal electricity consumption and no extra softener required).
Intelligent Reverse Osmosis controls water consumption, ensures a constant
product flow rate and optimal water quality. All prolong the lifetime of the RO
membrane and downstream purification technologies.
Pure water is stored in a polyethylene tank designed to minimize risks of
contamination during water storage. The vent filter provides effective
protection against airborne contaminants. The ASM, Automatic Sanitization
Module, prevents biofilm formation. Water can be either sourced through the
tank front valve or sent by a distribution pump to feed laboratory instruments.
The Progard® cartridge incorporates RFID technology. This provides instant
visibility on cartridge parameters and also ensures a safe & optimal
performance of your system, giving you full traceability.
With intuitive controls that
simplify operation, Elix® Essential is user-friendly, and provides just
the information required. A color-coded backlit LCD display clearly shows
message importance with system alert and alarm icons. Elix® Essential
system’s small footprint enables installation wherever it is convenient in your
laboratory —on or under the bench, or on a wall. A single replaceable
cartridge, an optional long-lasting UV lamp coupled with timely service alerts,
allows your system to run smoothly all year long while preserving your budget.
Millitrack® e-Solutions provide enhanced data management control, remote
access capabilities to the system dashboard, and long-term electronic archiving
for your Elix® Essential.
Resistivity at 25 °C |
Typically 10-15 MΩ•cm |
TOC |
< 30 ppb |
Bacteria * |
< 10 CFU/mL |
Production flow Rate |
3 L/h - Elix® Essential (UV)
3 |
* Elix® Essential UV versions
only |
Feed water quality |
Potable tap water as described in
US-EPA, EP and WHO norms |
Feed water connection |
½” Gaz M |
Feed water pressure* |
1 - 6 bar |
Feed water temperature |
5 - 35 °C |
* For pressure above 6 bar, a
pressure regulator needs to be installed upstream of the system |
Dimensions (H x W x D) |
470 x 268 x 339 mm (18.5 x 10.5 x
13.3 in) |
Operating weight |
17.2 - 18.5 kg (46.8 - 49.6 lb) |
Electric power supply voltage |
100 - 230 V +/- 10% |
Electric power supply frequency |
50 - 60 Hz +/- 10% |
Elix® Reference main unit data
connection |
Ethernet (RJ45) |