Synergy® systems are
designed for laboratories that already have access to purified water and need a
point-of-use system to supply ultrapure water.
With the Synergy® range of water purification systems, you benefit from a
choice of ultrapure water configurations. All systems produce Type 1 water
(18.2 MΩ•cm at 25 °C ultrapure water) on demand from pretreated water. For
applications sensitive to organic contaminants, a UV version of the Synergy® system
is also available. It incorporates a dual wavelength UV lamp for photooxidation
of organic compounds. A final purification step at the point-of-dispense
removes specific contaminants critical for your experiments. A range of
purification cartridges and Application-Paks is available for you to select
according to your requirements.
Simple handling & maintenance
·
Easy installation anywhere in your laboratory with its
compact design and optional remote dispenser. Either wall-mounted, on or under
the bench, choose the setup that is most convenient for your daily work.
·
Dispensing adapted to different size laboratory glassware at
an instant flow rate of 1.5 L/min.
·
Save time thanks to the volumetric function. It will
automatically deliver the water volume that you requested.
·
Easily insert your pack in one movement with the patented
locking mechanism.
Water Specifications
Synergy® ultrapure water (Type 1)
Resistivity at 25 °C |
18.2 MΩ•cm |
TOC1 (Synergy® system) |
≤ 10 ppb |
TOC1 (Synergy® UV
system) |
≤ 5 ppb |
Particulates (size > 0.22 μm)2 |
< 1 particulate/mL |
Bacteria (Synergy® system)2 |
< 0.1 CFU/mL |
Bacteria (Synergy® UV
system) 3 |
< 0.01 CFU/mL |
Pyrogens (endotoxins)3 |
< 0.001 EU/mL |
RNases3 |
< 0.01 ng/mL |
DNases3 |
< 4 pg/μL |
Flow Rate |
> 1.5 L/min |
1in
the appropriate test conditions |
Feed water
Feed water quality |
Elix®, RiOs™, distilled or DI
water with conductivity < 100 μS/cm and TOC < 50 ppb |
Feed water connection |
½” Gaz M |
Feed water pressure * |
0 - 0.3 bar |
Feed water temperature |
5 - 35 °C |
*
For pressure above 0.3 bar, a pressure regulator needs to be installed
upstream of the system; for pressures between 0 and -0.2 bar, the system will
operate, but product flow rate may be lower. |
System Specifications
Dimensions (H x W x D) |
540 x 290 x 380 mm (21.3 x 11.4 x
15 in) |
Operating weight (Synergy® systems) |
9.7 kg (21.4 lb) |
Operating weight (Synergy® UV
system) |
10.2 kg (22.5 lb) |
Operating weight (Remote
dispenser) |
2.68 kg (5.91 lb) |
Electrical power supply voltage |
100 - 250 V +/- 10 % |
Electrical power supply frequency |
50 - 60 Hz +/- 10 % |