Easy Programming and Monitoring:
·
Chamber and sample temperatures and operation status simultaneously
display on backlit LCD.
·
Chamber temperature is monitored by fixed-mount Type T thermocouple.
·
Sample temperature is monitored by Type T thermocouple with flexible cable
(included).
·
Pre-Cool step automatically begins temperature ramp when chamber and
sample temperatures are equilibrated within 1°C.
·
Audible indicators alert operator when hold time has elapsed.
·
Operation continues if power outage lasts less than 10 seconds;
audible/visual alarm sounds if outages last longer than 10 seconds.
·
Modifiable access code offers added security.
LN2 Freezing System Components:
·
Intuitive control panel and integrated thermal printer
·
Dual solenoid valves that increase LN2 injection volume for
precise temperature control and accelerated freezing for optimum cell viability
·
Specially designed air handling system and liquid nitrogen ring that work
in tandem for precise temperature control and uniformity
·
Overtemperature protection on heating element
Construction:
·
Type 304 stainless steel
·
Exterior powder-coat finish
·
Foamed in-place polyurethane insulation
·
Bulb-style gaskets for tight door seal; PTFE heat-breaks to prevent door
from freezing closed
·
Front- and top-access models designed for human IVF and easy “seeding”
Includes:
6′ (1.8m) flexible stainless-steel transfer
hose, 0.5″ 35° flare, thermal paper, 25- to 9-pin serial interface
cable/adapter, and 10′ (3m) RS-232 cable